Abstract
This method obeys Beers law in employed concentration range 2.4-12 µg/ml and 0.8-4 µg/ml for Pregabalin and Etoricoxib, respectively. The correlation coefficient of Etoricoxib was found to be 0.9993 and 0.9997, respectively. The retention time was 2.680 and 7.383 minutes. The percentage RSD for accuracy and precision was found to be less than 2%. The method was validated as per ICH guidelines for its selectivity, specificity, system suitability, linearity, range, precision, accuracy, LOD, LOQ, robustness, assay. The method was successfully employed for routine quality control analysis of Pregabalin and Etoricoxib in pharmaceutical formulation